The Settings page is organised into sections down the left-hand side. Nothing takes effect until you press Save Settings.

System

Field What it is
Testing Lab Code/ID A unique identifier for the laboratory, e.g. LAB001. It travels with every result.
Testing Lab Name The laboratory's name, as it should appear on results.
Display Date Format How dates are shown on screen: 16-Sep-2026 (the default), 16-09-2026, 16/09/2026, 16.09.2026, 09/16/2026 or 2026-09-16. Each choice is shown with today's date. It changes only the display. Stored results, raw data and what the LIS receives keep their own format. Hover over a date in the results to see the stored value.
Time Zone The zone used to record when a result or transmission is received, shown as Received On. It starts as this computer's zone. Type part of a name or an offset to search, such as Kampala or +03:00. Times sent by the instrument, such as when a test was analysed, are stored as sent. Results already stored keep their time. Before 4.8.0 the received time came from the database: UTC in SQLite and the server's own zone in MySQL.
Auto-connect on startup Yes skips the login screen and connects every instrument when the application opens. No shows the login screen and leaves connecting to you.

The SQLite Database Path is shown rather than asked for: it is where results are stored on this machine, and the tool decides it.

MySQL (optional)

Configure this if your LIS reads results from MySQL.

Field What it is
MySQL Host Database server address, e.g. 127.0.0.1
MySQL Port Default 3306
Database Name e.g. interfacing
Database User MySQL username
Database Password MySQL password

Test Connection proves the database is reachable before you save a configuration that depends on it.

MySQL is optional

Without it, results are still stored locally in SQLite. What you lose is the shared table an LIS on another machine can read.

Instruments

+ Add Instrument for each analyzer this tool will talk to.

Connection

Field What it is
Connection Mode TCP Server — the analyzer connects to this tool. TCP Client — this tool connects to the analyzer. The analyzer's own configuration decides which one you need.
Communication Protocol ASTM, ASTM (with checksum), or HL7.
IP Address In server mode, the address on this machine to listen on. In client mode, the analyzer's address.
Port Number 1–65535, matching what the analyzer is configured for.

Checksum or not is the analyzer's decision, not a preference

An analyzer configured to send checksums and a tool configured not to expect them will still appear to work, and will still store results. What is lost is the check that a frame arrived intact, and the ability to ask for it again when it did not. Set this to what the analyzer is actually doing. If you are unsure, the console log shows which one it is reading.

The instrument itself

Field What it is
Analyzer Type The model. Roche cobas Taqman, 4800, 5800, 6800/8800; Abbott m2000, Alinity m; Cepheid GeneXpert; or one of the generic ASTM and HL7 choices.
Instrument Name/Code The name the LIS knows this analyzer by. It is what results are mapped by, so it has to match what the LIS expects.
Display Order The order the instruments appear in on the console.

Each instrument needs a unique name and a unique address-and-port combination.

Why the analyzer type matters

Two analyzers can both speak ASTM and still disagree about where the sample identifier lives or what an empty result field means. The analyzer type is how the tool knows which of those dialects it is listening to. Choosing "Other" for a model that is in the list will store results, but some fields may come through empty.

Result rules

Result rules tell the tool to store a different value when an instrument sends a particular result. Use them when your LIS expects a result written in a way the instrument does not send it. Each instrument has its own rules.

Field What it is
Is exactly / Contains Whether the whole result must match, or only part of it. Spaces around the result are ignored.
Result The text the instrument sends, e.g. > Titer max.
Store instead The value to store. It replaces the whole result, exactly as typed.
Ignore case Match NOT DETECTED and Not Detected alike.

The first rule that matches is used. A result that no rule matches is stored exactly as the instrument sent it. Failed and Incomplete are never replaced, so no rule can make a failed run look like a result.

  • A rule with either text left empty is ignored.
  • Contains also matches inside numbers: a rule for 20 matches 1200.
  • Rules stay with the instrument if you change its protocol. Review them when you do.

  • Duplicate copies one rule, to change it slightly.

  • Copy from another instrument lists the instruments that have rules. Choose one to add its rules to this instrument. Rules this instrument already has are skipped.

An instrument with no rules shows a single line under Result rules, with Add rule. When another instrument has rules, the line also shows Copy from another instrument.

The result as read before any rule is always kept beside the stored one, and the raw data never changes. After changing a rule, reprocess older raw data to store earlier results the new way. If the rule changes a result, reprocessing stores a new result beside the old one. If the rule leaves a result unchanged, reprocessing does not store it again. See raw data and recovery.

Rules you may already have

Earlier versions always stored > Titer max as > 10000000 and <20 as < 20 for HL7 instruments. Those are now the starting rules of every HL7 instrument configured before this version, so nothing changes until you edit or remove them. A new instrument starts with no rules.

Check with your LIS administrator before adding a rule. The LIS reads the value you store, not the one the instrument sent.

LIS API (optional)

If your LIS offers an API, the tool can fetch the instrument names it expects, so the names on both sides match.

Field What it is
Base URL e.g. https://lis.example.org
Auth Type None, Bearer Token, Basic Auth, or API Key
Fetch Instruments Endpoint e.g. /api/v1.1/instruments?labId=XYZ

Fetch Instruments tests the connection and retrieves the names. You can always type them by hand instead.

Result forwarding (optional)

Result forwarding sends every new result to one more system over HTTP, for example an integration engine such as Open Integration Engine or Mirth Connect. It does not change MySQL or InteLIS delivery. This section has its own Save forwarding button. Save Settings does not save it.

Field What it is
Forward results to this receiver Turns forwarding on or off.
Receiver URL The address results are sent to, e.g. http://localhost:8081/results.
Authentication None, Bearer token, Basic (username and password), or API key (X-API-Key header).
Username For Basic authentication only.
Bearer token, Password or API key The secret for the chosen authentication. After saving, leave it blank to keep the saved secret.

Send test sends an empty test request with the values in the form. It sends no results.

The first time you save this section, results already stored on this computer are marked as not to be sent. This happens whether forwarding is on or off. Results received after that first save are sent while forwarding is on. While it is off, they are kept and sent once you turn it on. A result stays queued until the receiver accepts it. Waiting to send shows how many are queued.

If you change the receiver URL to a different server, enter the secret again. A saved secret is only ever sent to the server it was saved for.

A URL that uses plain http:// to another computer shows a warning. Results and credentials cross the network unencrypted on that URL.

Settings exports and backups do not include result forwarding. After restoring on another computer, set it up again. The result webhook reference describes the request a receiver gets.

Connecting to InteLIS

If your laboratory uses InteLIS, the Connection Code from your facility page configures the laboratory, its instruments and the credentials in one step, rather than filling in the sections above. Codes are single-use and expire.