The console is where you connect instruments and watch results arrive.

Instrument connections

Each configured instrument is a tab across the top, showing:

  • Connection status — connected, or not
  • Connection details — model, mode, address, protocol
  • Connection logs — what is happening on the wire, as it happens

To connect: if Auto-connect is on, the tool connects everything at startup. If it is off, sign in and press Connect on each tab.

Results only arrive while an instrument shows as connected

An analyzer that finishes a run while the tool is disconnected does not normally queue the result and send it later. Most will report the transmission as failed and leave the result to be re-sent by hand from the analyzer's own screen.

Connection logs

The log is the first place to look when something is not working, and it is written to be read by someone who is not a programmer. Lines you will see regularly:

Line What it means
Server bound and listening on <address>:<port> The tool is ready and waiting for the analyzer to connect.
Client <address> has connected The analyzer opened a connection.
Receiving.... followed by record text A transmission is arriving.
Sending ACK The tool acknowledged what it received, as the protocol requires.
Successfully saved result : <sample> A result was stored.
Client <address> disconnected normally The analyzer closed the connection, which it does between runs.

Received results

Below the tabs, every result the tool has stored:

Column What it is
Instrument Which analyzer sent it
Sample ID The sample identifier, as the analyzer sent it
Result The result, as the analyzer sent it
Unit The unit, as the analyzer sent it
Test Type The assay
Tested By The operator recorded by the analyzer
Tested On When the analyzer ran the test
Received On When this tool received it
Sync Status Pending, Synced or Failed

You can search by instrument, sample ID or test type, re-sync results that failed, and refresh the list.

Why a result can look odd

Results are stored exactly as the analyzer reported them. An analyzer set to French sends NON DÉTECTÉ and writes numbers with a decimal comma (1420403,41); an Abbott m2000 sends < 40; a Roche 6800 sends a mantissa with a power-of-ten unit. None of that is corrected on the way in, because correcting it would mean guessing, and a guess about a viral load is not something a laboratory can check afterwards. If the LIS needs another form, that conversion belongs in the LIS.

Quick stats

Two timestamps at the bottom:

  • Last Results Synced to LIS — when the LIS last took results
  • Last Result From Instrument — when an analyzer last sent one

They answer different questions. Results arriving but not syncing is a problem between the tool and the LIS; nothing arriving at all is a problem between the analyzer and the tool.

Raw data

View Raw Data shows the transmissions themselves, as received. It is the record of what the analyzer actually said, kept so that a result can be re-derived if the tool ever read one wrongly — see raw data and recovery. When reporting a problem with a result, this is the useful thing to send.

Show original data, the file icon at the end of a result's row, shows two things: the result's own records, and the whole transmission they arrived in.